OXYTRANS TR – Oxygen (Optical; In-line)
In-line O₂ Measurement in Liquids and Gases
OXYTRANS TR is an optical oxygen sensor for continuous O₂ measurement in deaerated water, wort, beer, beverages, liquid food products, CO₂ and inert gases.
Measuring Principle
Indicator molecules in the optical window are excited by blue-green light. Oxygen transfers the excitation energy and changes the detected signal. As the O₂ concentration increases, the signal decreases.
Applications
- Deaerated Water
- Wort
- Beer
- Beverages
- Liquid Food Products
- CO₂
- Inert Gases
- Product Line
OXYTRANS TR is designed for applications requiring maximum sensitivity and highest accuracy.
Technical Data
Measuring Ranges, Liquid Phase: I) 1 ppb - 2 ppm
II) 30 ppb - 35 ppm
Accuracies, Liquid Phase: I) ± 1 ppb
II) ± 30 ppb
Measuring Ranges, Gas Phase: I) 0 - 4,2 % O₂
II) 0 - 50 % O₂
III) 0 - 0,02 % O₂
Accuracies, Gas Phase: I) ± 0,002 % O₂
II) ± 0,03 % O₂
III) ± 0,0002 % O₂
Range III Restriction: Gas phase and standard temperature range only
Response Time / Operating Pressure: t90 ≤ 10 s / max. 12 bar
Operating Temperature: -5 to +55 °C (Pt100); optional +98 °C
Material: 1.4404 / AISI 316L, silicone (FDA), PTFE (FDA)
Input: 1 x analog (4 - 20 mA) & 3 x digital (24 VDC)
Output / Fieldbus: 2 x analog (4 - 20 mA) & 3 x digital (24 VDC)
optional: PROFIBUS DP & PROFINET
Enclosure Rating / Supply: IP65 / 24 VDC
Optical Measurement: No replacement of electrolyte or membrane.
Short Response Time: t90 ≤ 10 s.
In-line Integration: Continuous measurement in the process.
Modular Design & Extensions
DeGaS-Cold – Column Water Deaeration (Cold)
DeGaS-Hot – Column Water Deaeration (Hot)
DGS – Membrane Water Deaeration