NITROTEC TC is a thermal-conductivity sensor for continuous measurement of dissolved and gaseous nitrogen (N₂). It is suitable for in-line measurements from trace concentrations to saturation and for applications in beverage production, power and steam, and pharmaceuticals.
Applications
- Beer and beverage production, particularly nitrogenated products
- Power & Steam
- Pharmaceutical processes
Measuring Principle
N₂ diffuses through a gas-permeable membrane into the measuring chamber, which is continuously flushed with CO₂ as the carrier gas. A thermal-conductivity detector measures the change in thermal conductivity, and the dissolved concentration is derived from the N₂ partial pressure.
Technical Data
- Dissolved N₂: 0–350 ppm
- N₂ partial pressure: 0–20 bar
- Cycle time: 15 s
- Linear flow rate: 150 cm/s
- Purge gas: CO₂
- Accuracy: ±1 % of reading*
- Temperature Compensation: 0 to +45 °C
- CIP/SIP: up to 120 °C
- Process pressure: 0–20 bar
- Wetted materials: 316L stainless steel, PFA
- Enclosure Rating: IP65
- Weight: 0,6 kg
- Sensor cable: 3–100 m
Extension
Optionally with CENTEC TC ANALYZER as the evaluation unit for display, alarm functions and signal integration.
* 1 % of the reading or the minimum value, whichever is greater.
Evaluation unit / Analyzer
NITROTEC TC is optionally available with the CENTEC TC ANALYZER as the evaluation unit. The analyzer displays concentration and temperature and provides alarm and signal integration into the process control system.Downloads (PDF, English)