CENTEC TC ANALYZER
An evaluation unit for the Centec thermal-conductivity sensor family. It complements NITROTEC TC and HYDROTEC TC and continuously measures N₂ or H₂ concentration and medium temperature in liquid and gas phases. Parameters can be adjusted to the measuring conditions and operating requirements. The stainless-steel enclosure is designed for continuous process measurements in demanding applications.
Operation and Integration
- Continuous LCD display of concentration, medium temperature and real-time clock.
- Quick configuration using function keys and on-screen menus.
- PIN-code-protected access levels and software-guided sensor calibration.
- Configurable concentration alarms and integration via digital and analog outputs.
Technical Data
Power Supply: 85-230 VAC, 50/60 Hz
Power consumption: Max 20 W
Operating conditions: -5 °C to 50 °C
Enclosure: Stainless steel, IP65 / Nema4X
Weight: 3.0 kg
Dimensions: 230 × 180 × 75 mm
Dissolved range: N₂ 0-350 ppm · H₂ 0-2 ppm
Gaseous range: N₂ 0-20 bar · H₂ 0-1.5 bar
Dissolved display units: ppm, ppb, mg/l
Dissolved resolution: 0.01 ppb
Gaseous display units: %-ppm, kPa
Gaseous resolution: 0.01 ppm
Displayed data: Temperature, measured value, real-time clock, membrane type
Alarm outputs: High/low or high/high-high relays, resistive load 125 VAC / 20 VDC
Thermal cut-off: Factory setting +40 °C / 140 °F, adjustable via software
Digital outputs: RS 232 / RS 485
Analog outputs: 0-20 mA / 4-20 mA
Calibration: Single-point air and direct calibration
Compatible sensors: NITROTEC TC · HYDROTEC TC
Downloads (PDF, English)

